Message Displayed | Root Cause | Suggestion |
ERR! 1 | ■ EEPROM(24LC1025)self-test error. | Replace EEPROM |
ERR! 2 | Audio board self-test error | Change a new audio board ( sound board) |
ERR! 3 | Drop sensor self-test error | Replace pressure sensor |
ERR! 4 | Pressure sensor self-test error | Replace pressure sensor |
ERR! 5 | Air sensor self-test error | Replace air sensor module |
ERR! 6 | Pressure calibration error. | Pressure accuracy is not calibrated, recalibrate it accordingly |
ERR! 8 | Free flow alarm. The administration set is not installed properly. | Check and reinstall the administration set. |
ERR! 9 | Key stuck during infusion/pause alarm. Some of the keys is broken. | Check and replace the keyboard |
ERR! 20 | No battery alarm. No battery in the device. The battery is dead. The battery is badly connected. | Check battery connection wires. Check battery connector. Check battery |